100MHz Analog Oscilloscope 2245A

Tektronix; Portland, OR

  • Anno
  • 1986–1989
  • Categoria
  • Strumento da laboratorio
  • Radiomuseum.org ID
  • 246391

 Specifiche tecniche

  • Numero di tubi
  • 1
  • Valvole
  • Numero di transistor
  • A semiconduttori.
  • Semiconduttori
  • Gamme d'onda
  • - senza
  • Tensioni di funzionamento
  • Alimentazione a corrente alternata (CA) / 90-250 Volt
  • Altoparlante
  • - - Nessuna uscita audio.
  • Materiali
  • Mobile di metallo
  • Radiomuseum.org
  • Modello: 100MHz Analog Oscilloscope 2245A - Tektronix; Portland, OR
  • Forma
  • Soprammobile con qualsiasi forma (non saputo).
  • Dimensioni (LxAxP)
  • 362 x 164 x 445 mm / 14.3 x 6.5 x 17.5 inch
  • Annotazioni
  • VERTICAL SYSTEM (4 channels)
    Bandwidth (-3dB) and Rise Time: 100 MHz and 3.5 ns (-10°C to +35°C); 90 MHz and 3.9 ns (2 mV/div or +35°C to +55°C). Bandwidth limit: 20 MHz.

    Deflection Factor and Accuracy: CH 1&2: 2 mV/ div to 5 V/div; CH 3&4: 0.1 V/div and 0.5 V/div; all at ±2% (±3% outside +15°C to +35°C). CH 1&2 variable at least 2.5:1

    Vertical Operating Modes: CH 1, 2, 3, 4, CH 2 INVERT, ADD, ALT, CHOP (625 kHz).

    CMRR: At least 10:1 at 50 MHz.
    Input R and C: 1 MΩ, 20 pF.
    Max Input Voltage: 400 V (dc +peak ac) or 800 Vp-p.
    Channel Isolation: 50:1 at 100 MHz.

    HORIZONTAL SYSTEM
    Sweep Speeds: A Time Base: 0.5 s/div to 20 ns/div; B Time Base: 5 ms/div to 20 ns/div (X10 MAG to 2 ns/div A and B)
    Accuracy: ±2%; Magnified ±3% (degrade by 1% outside +15°C to +35°C).
    Horizontal Operating Modes: A, ALT, B, X-Y.
    Delay Jitter: 20,000:1.
    Delay Accuracy: ±0.5% + 5% of one div +25 ns.

    TRIGGER SYSTEM
    Trigger Sensitivity (A and B)- dc: 0.35 div to 25 MHz, 1.0 div at 150 MHz.  Noise Reject: 1.4 div to 25 MHz, 2.2 div at 100 MHz.
    HF Reject: attenuates above 70 kHz. LF Reject: attenuates below 50 kHz. ac: Same as dc, attenuates below 25 Hz. TV Line, TV Field: 0.5 div of composite sync for stable display.

    Trigger Operating Modes: A mode: AUTO LEVEL, AUTO, NORM, TV LINE, TV FIELD, SINGLE SEG. B mode: RUNS AFTER DELAY, AUTO LEVEL, NORM, TV LINE (from A source).
    Trigger Source (A and B): VERT, CH 1, 2, 3, 4, LINE.
    Variable Holdoff: At least 10:1.

    X-Y OPERATION
    Deflection Factors: Same as vertical system.
    X-Y Operating Modes: X: CH 1; Y: CH 1, 2, 3, 4, and ADD
    Bandwidth:  X-Axis: 3 MHz; Y-Axis same as vertical System.
    Phase Difference: ±3° from dc to 50 kHz.

    CRT SYSTEM
    Display: 8 cm x 10 cm, 16.5 kV nominal voltage.
    Controls: BEAM FIND, FOCUS, A and B READOUT INTENSITY, TRACE
    ROTATION and SCALE ILLUM.
    Z-Axis: 3.8 volts causes noticeable modulation. Usable to 10 MHz.
     

  • Peso netto
  • 7.9 kg / 17 lb 6.4 oz (17.401 lb)
  • Bibliografia
  • -- Original prospect or advert
  • Letteratura / Schemi (1)
  • Funkschau (18/1986, S. 68 / Neuvorstellung)
  • Autore
  • Modello inviato da Pius Steiner. Utilizzare "Proponi modifica" per inviare ulteriori dati.

 Collezioni | Musei | Letteratura

Collezioni

Il modello 100MHz Analog Oscilloscope fa parte delle collezioni dei seguenti membri.

 Forum