IET Output Power Meter 1840-A
General Radio Company; Cambridge (MA)
- Land
- USA
- Hersteller / Marke
- General Radio Company; Cambridge (MA)
- Jahr
- 1962
- Kategorie
- Service- oder Labor-Ausrüstung
- Radiomuseum.org ID
- 320468
-
- anderer Name: GenRad
Klicken Sie auf den Schaltplanausschnitt, um diesen kostenlos als Dokument anzufordern.
- Wellenbereiche
- - ohne
- Betriebsart / Volt
- Keine Stromversorgung
- Lautsprecher
- - - Kein Ausgang für Schallwiedergabe.
- Material
- Metallausführung
- von Radiomuseum.org
- Modell: IET Output Power Meter 1840-A - General Radio Company;
- Form
- Reisegerät > 20 cm (netzunabhängig betreibbar)
- Abmessungen (BHT)
- 305 x 90 x 170 mm / 12 x 3.5 x 6.7 inch
- Bemerkung
-
The IET 1840-A measures audio-frequency power into any desired load. This model formerly was manufactured by GenRad (General Radio).
Its important uses include the measurement of:
- *Power output of oscillators, amplifiers, preamplifiers, transformers, transducers, and low-frequency lines.
- *Output impedance, by adjustment of the load, to yield maximum power indication.
- *Frequency-response characteristics of amplifiers, transformers, and other audio-frequency devices.
This instrument basically is a multi-tapped audio-frequency transformer with a fixed secondary load. Its two front-panel switches connect eight identical primary windings and six secondary taps in various combinations to provide a total of 48 different primary impedances.
The maximum power rating can be extended for any given impedance with the use of simple T-network attenuator, design data for which are supplied with the instrument.
Power Range:
0.1 mW to 20 watts in five ranges, 40 Hz to 20 kHz; maximum rating is reduced by up to 50% (at 25 Hz), depending on impedance selected.
An auxiliary dB scale on the meter reads from -15 to +3 dB, referred to 1 mW.
Impedance Range:
0.6 Ω to 32 kΩ in two ranges; yielding 48 individual impedances spaced approximately 2 apart.
Accuracy
Impedance:
at 1 kHz, ±6% max, -0.5% avg;
70 Hz to 2.5 kHz, ±7% above 10 kW;
70 Hz to 5 kHz, ±7% below 10 kW;
at 20 Hz, -15% max, -8% avg;
at 20 kHz, ±50% max, ±12% avg
When impedance is specified, accuracy applies at full-scale on all ranges and impedances at 1 kHz;
when frequency ranges are specified, accuracy is at full-scale at ranges ≥20 mW at the specified impedances.
When a frequency limit is specified, accuracy is at full-scale at all impedances
Power:
at 1 kHz, ±0.3 dB;
50 Hz to 6 kHz, ±0.5 dB;
30 Hz to 10 kHz, ±1 dB;
at 20 Hz, -1.5 dB max, -1 dB avg;
at 20 kHz, -5 dB max, ±1.5 dB avg.
Waveform Error:
Meter will indicate true rms with as much as 20% second and third harmonics present in the input signal