Coaxial Frequency Meter 537A

Hewlett-Packard, (HP); Palo Alto, CA

  • Anno
  • 1972
  • Categoria
  • Strumento da laboratorio
  • Radiomuseum.org ID
  • 338595

 Specifiche tecniche

  • Principio generale
  • RF-Circuit or wave trap (separate, can also be an antenna with amp.)
  • Gamme d'onda
  • Gamme d'onda nelle note.
  • Tensioni di funzionamento
  • Alimentazione non richiesta
  • Altoparlante
  • - - Nessuna uscita audio.
  • Materiali
  • Mobile di metallo
  • Radiomuseum.org
  • Modello: Coaxial Frequency Meter 537A - Hewlett-Packard, HP; Palo Alto
  • Forma
  • Diverse forme (descritte nelle note).
  • Dimensioni (LxAxP)
  • 118 x 146 x 89 mm / 4.6 x 5.7 x 3.5 inch
  • Annotazioni
  • The Hewlett Packard, 537A coaxial frequency meter is used for measuring frequency in the 0.96 to 12.4 GHz range.

    Accuracy is obtained from the spiral scale.

    The meter contains a coaxial line coupled to a quarter-wavelength coaxial resonant cavity loaded with capacitance at the bottom of the band. At resonance, power is absorbed by the cavity, producing a dip in the coaxial line power.
    The frequency at which this dip occurs is read directly from the dial in gigahertz.

    Tuning is accomplished by moving a piston. A precision lead screw, which is spring-loaded to prevent backlash, is used to position the piston in the cavity.

    Due to the design of the cavity, its electrical length is extended at lower frequencies, and there are no spurious modes.

    Specifications

    Frequency range;   3.7 to 12.4 GHz
    Reflection coefficient off-resonance: 0.33 (2.0 SWR, 9.5 dB return loss)
    Calibration increments: 10 MHz
    Dip at resonance:

    1 dB (minimum)

    Dial accuracy:   ±0.1%
    Overall Accuracy:

    ±0.17% which includes an allowance of ±0.02% for 0 to 100% relative humidity, ±0.0016 per °C from 13 to 33 °C & 0.03% backlash.

     

  • Peso netto
  • 3.5 lb (3 lb 8 oz) / 1.589 kg
  • Bibliografia
  • - - Manufacturers Literature (536A / 537A Coaxial Frequency Meter Operating Note. No. 00536-90001)
  • Autore
  • Modello inviato da Gary Cowans. Utilizzare "Proponi modifica" per inviare ulteriori dati.