Spectrum Analyzer System 141T / 8552B / 8554B

Hewlett-Packard, (HP); Palo Alto, CA

  • Année
  • 1969 ?
  • Catégorie
  • Appareils de mesure et de dépannage (matériel de labo)
  • Radiomuseum.org ID
  • 190235

 Spécifications techniques

  • No. de tubes
  • 1
  • Lampes / Tubes
  • No. de transistors
  • Semi-conducteurs présents
  • Semi-conducteurs
  • Gammes d'ondes
  • Bandes en notes
  • Tension / type courant
  • Alimentation Courant Alternatif (CA) / 115; 230 Volt
  • Haut-parleur
  • - - Pas de sortie basse fréquence
  • Matière
  • Boitier métallique
  • De Radiomuseum.org
  • Modèle: Spectrum Analyzer System 141T / 8552B / 8554B - Hewlett-Packard, HP; Palo Alto
  • Forme
  • Modèle de table profil bas (grand modèle).
  • Dimensions (LHP)
  • 425 x 221 x 416 mm / 16.7 x 8.7 x 16.4 inch
  • Remarques
  • This HP Spectrum Analyzer System consists of the Display System 141T (Variable Persistance Display), the High Resolution IF Section 8552B and the plug-in 8554B (Spectrum Analyzer RF Section). It is part of Hewlett Packard's high performance plug-in spectrum analyzer family which allows frequency domain measurements from 20 Hz to 40 GHz.
    Some specifications of this combination:
    Frequency range 100 kHz to 1250 MHz, flatness +/- 1 dB.
    Resolution bandwidth: IF bandwidths of 0.1 to 100 kHz provided in a 1,3 sequence.
    Dynamic range: Average noise level <-102 dBm with 10 kHz IF bandwidth
    RF input impedance: 50 ohm, max. input level + 13 dBm (1.4 Vac peak).
    Scan time: 16 internal scan rates from 0.1 ms/div to 10 sec/div in a 1, 2, 5 sequence, and manual scan.
    The HP 141T/8552B can also be equipped with the RF plug-ins 8553B (1 kHz to 110 MHz), 8556A (20 Hz to 300 kHz, has internal tracking generator), 8555A (10 MHz to 40 GHz). Also available: Tracking Generators for each frequency band, pre-amplifier 8447 series, matching transformer 3-500 MHz 11694A, external mixer 11517A, limiter 0.1-12.4 GHz 11693A, active probe 0.1-500 MHz 1121A. 1980 still in catalog listed.
  • Poids net
  • 26.8 kg / 59 lb 0.5 oz (59.031 lb)
  • Littérature