Electronic Counter TF 2401
Marconi Scientific Instrument Co. LTD.,The; London
- Pays
- Royaume Uni
- Fabricant / Marque
- Marconi Scientific Instrument Co. LTD.,The; London
- Année
- 1964 ?
- Catégorie
- Appareils de mesure et de dépannage (matériel de labo)
- Radiomuseum.org ID
- 339672
Cliquez sur la vignette du schéma pour le demander en tant que document gratuit.
- No. de transistors
- Semi-conducteurs présents
- Semi-conducteurs
- Gammes d'ondes
- Bandes en notes
- Tension / type courant
- Alimentation Courant Alternatif (CA)
- Haut-parleur
- - - Pas de sortie basse fréquence
- Matière
- Boitier métallique
- De Radiomuseum.org
- Modèle: Electronic Counter TF 2401 - Marconi Scientific Instrument
- Forme
- Modèle de table profil bas (grand modèle).
- Dimensions (LHP)
- 0 x 5.25 x 0 inch / 0 x 133 x 0 mm
- Remarques
-
The Marconi TF 2401 counter is a basic instrument into which plugs a frequency range unit and a function unit.
The plug-in units at present available are a 50 MHz frequency range unit (TM 7557), a counter-timer function unit, and a 500 MHz converter unit (TM 7558).
When used with these units the instrument performs the following basic measurements: Frequency, period, multiple periods, time interval, frequency ratio, and many other instructions.
Features include an 8-digit read-out with memory.Abridged Specifications
Basic Counter TF 2401
DISPLAY: 8 digits in line, with memory.
GATE TIMES (frequency measurement): 1 µsec to 100 sec in decade steps, instrument reads in c/s, kc/s, or Mc/s, with automatic decimal point indication.STANDARD FREQUENCY STABILITY:
+ 2 parts in 10' short term.
± 1 part in 103 per day.50 Mc/s Range Unit TM 7557
(Input Channel A)RANGE: 1 C/S to 50 Mc/s.
SENSITIVITY: 0.1 V RMS., 10 c/s to 50 Mc/s.
INPUT IMPEDANCE: Approximately 500 kΩ shunted by 25 pF.Counter Timer Unit TM 7558
(Input Channels B and C)RANGE: 0 to 10 MC/S.
SENSITIVITY: O.1 V RMS. 1 c/s to 10 Mc/s.
INPUT IMPEDANCE: Approximately 1MΩ shunted by 30 pF.PERIOD AND TIME INTERVAL MEASUREMENTS:
TIMING SIGNALS: 1 c/s to 10 Mc/s in decade steps.
The instrument reads in sec, msec, or µsec, with automatic decimal point indication.MULTIPLE PERIOD MEASUREMENT:
TIMING SIGNALS: 1 Mc/s or 10 Mc/s. The instrument reads in µsec with decimal point indication.
MULTIPLYING FACTOR: (Multiple periods and multiple frequency ratio measurements) 10 to 108 in decade steps.
- Littérature
- “Electronic Engineering”, volume 36, June 1964.
- Schémathèque (1)
- - - Manufacturers Literature (Marconi Instrumentation Vol. 9 No. 6, June 1964, Pages 132-138)
- Auteur
- Modèle crée par Pier Antonio Aluffi. Voir les propositions de modification pour les contributeurs supplémentaires.
- D'autres Modèles
-
Vous pourrez trouver sous ce lien 36 modèles d'appareils, 36 avec des images et 4 avec des schémas.
Tous les appareils de Marconi Scientific Instrument Co. LTD.,The; London