Pulse Generator 114

Tektronix; Portland, OR

  • Année
  • 1965
  • Catégorie
  • Appareils de mesure et de dépannage (matériel de labo)
  • Radiomuseum.org ID
  • 250343

 Spécifications techniques

  • No. de transistors
  • 19
  • Semi-conducteurs
  • Gammes d'ondes
  • - sans
  • Tension / type courant
  • Alimentation Courant Alternatif (CA) / 115; 230 Volt
  • Haut-parleur
  • - - Pas de sortie basse fréquence
  • Matière
  • Boitier métallique
  • De Radiomuseum.org
  • Modèle: Pulse Generator 114 - Tektronix; Portland, OR
  • Forme
  • Modèle de table générique
  • Dimensions (LHP)
  • 9 x 6 x 12.5 inch / 229 x 152 x 318 mm
  • Remarques
  • The Tektronix Type 114  is a general-purpose generator designed for laboratory and production  test facilities. It utilizes solid-state circuitry throughout. Three switches are used for selecting pulse period, width and amplitude. Each switch has a variable control  for use over the selected range. A front-panel lamp indicates when the pulse width exceeds the maximum allowed by the period. The width selector switch features a square-wave position providing symmetrical square-wave output at all pulse periods.

    • Rise and Fall time: 10 nsec or less
    • Pulse Aberrations: 5% or less (at maximum amplitude)
    • Polarity: Positive or negative
    • Pulse Period: 5 switches position from 1 µsec to 10 msec, plus external trigger position; continuously variable from 1 µsec to 100 msec. Accuracy: +/-3% with variable control in calibrated position. Jitter: less than 0.05%
    • Pulse width: 5 switched positions from 100 nsec to 1 msec, plus square wave mode; continuously variable from 100 nsec to 10 msec. Accuracy: +/-3% with variable control in calibrated position. Jitter: less than 0.05%. Front panel lamp indicates when pulse width exceeds pulse period. Square wave output: symmetrical square-wave output at all pulse period settings.
    • Pulse Amplitude: 4 switched positions: -1 to -3 V, -3 to -10 V, +1 to +3 V, +3 to +10 V. Variable within each range. Maximum output into a 50-Ω load is 10 V, 16 V into 1000-Ω load.
  • Poids net
  • 10 lb (10 lb 0 oz) / 4.540 kg
  • Littérature
  • - - Manufacturers Literature (Tektronix Catalog 24, Mid 1965 to Mid 1966, p.209)
  • Auteur
  • Modèle crée par Bernhard Nagel. Voir les propositions de modification pour les contributeurs supplémentaires.