Pulse Generator 114
Tektronix; Portland, OR
- Pays
- Etats-Unis
- Fabricant / Marque
- Tektronix; Portland, OR
- Année
- 1965
- Catégorie
- Appareils de mesure et de dépannage (matériel de labo)
- Radiomuseum.org ID
- 250343
- No. de transistors
- 19
- Semi-conducteurs
- Gammes d'ondes
- - sans
- Tension / type courant
- Alimentation Courant Alternatif (CA) / 115; 230 Volt
- Haut-parleur
- - - Pas de sortie basse fréquence
- Matière
- Boitier métallique
- De Radiomuseum.org
- Modèle: Pulse Generator 114 - Tektronix; Portland, OR
- Forme
- Modèle de table générique
- Dimensions (LHP)
- 9 x 6 x 12.5 inch / 229 x 152 x 318 mm
- Remarques
-
The Tektronix Type 114 is a general-purpose generator designed for laboratory and production test facilities. It utilizes solid-state circuitry throughout. Three switches are used for selecting pulse period, width and amplitude. Each switch has a variable control for use over the selected range. A front-panel lamp indicates when the pulse width exceeds the maximum allowed by the period. The width selector switch features a square-wave position providing symmetrical square-wave output at all pulse periods.
- Rise and Fall time: 10 nsec or less
- Pulse Aberrations: 5% or less (at maximum amplitude)
- Polarity: Positive or negative
- Pulse Period: 5 switches position from 1 µsec to 10 msec, plus external trigger position; continuously variable from 1 µsec to 100 msec. Accuracy: +/-3% with variable control in calibrated position. Jitter: less than 0.05%
- Pulse width: 5 switched positions from 100 nsec to 1 msec, plus square wave mode; continuously variable from 100 nsec to 10 msec. Accuracy: +/-3% with variable control in calibrated position. Jitter: less than 0.05%. Front panel lamp indicates when pulse width exceeds pulse period. Square wave output: symmetrical square-wave output at all pulse period settings.
- Pulse Amplitude: 4 switched positions: -1 to -3 V, -3 to -10 V, +1 to +3 V, +3 to +10 V. Variable within each range. Maximum output into a 50-Ω load is 10 V, 16 V into 1000-Ω load.
- Poids net
- 10 lb (10 lb 0 oz) / 4.540 kg
- Littérature
- - - Manufacturers Literature (Tektronix Catalog 24, Mid 1965 to Mid 1966, p.209)
- Auteur
- Modèle crée par Bernhard Nagel. Voir les propositions de modification pour les contributeurs supplémentaires.
- D'autres Modèles
-
Vous pourrez trouver sous ce lien 386 modèles d'appareils, 374 avec des images et 92 avec des schémas.
Tous les appareils de Tektronix; Portland, OR
Collections
Le modèle Pulse Generator fait partie des collections des membres suivants.