Time Base 3B4
Tektronix; Portland, OR
- Produttore / Marca
- Tektronix; Portland, OR
- Anno
- 1964
- Categoria
- Strumento da laboratorio
- Radiomuseum.org ID
- 231766
- Numero di tubi
- 4
- Numero di transistor
- 20
- Semiconduttori
- Gamme d'onda
- - senza
- Tensioni di funzionamento
- Fornita mediante altra unità o unità principale.
- Altoparlante
- - - Nessuna uscita audio.
- Materiali
- Mobile di metallo
- Radiomuseum.org
- Modello: Time Base 3B4 - Tektronix; Portland, OR
- Forma
- Chassis o in scatola da montaggio
- Dimensioni (LxAxP)
- 4.25 x 6.5 x 14.5 inch / 108 x 165 x 368 mm
- Annotazioni
-
The Type 3B4 Time Base plug-in unit is designed to be used primarily with the Tektronix Type 561A Oscilloscope system; however, It can also be used with the other 560-Series systems that use 3B-Series plug-in units such as the Type 564 and Type 567. As part of an oscilloscope system, the Type 3B4 is used as an accurate time base generator or as a calibrated amplifier for externally generated deflection signals.
When used with the 567 or RM567, the 3B4 provides a time-base, but does not activate the digital readout circuitry. In the 565 or RM565, the 3B4 provides a vertical time-base for raster applications but does not provide retrace blanking.Sweep
TIME/DIV Switch
0.2 µsec/div to 5 sec/div in 23 steps, 1, 2, 5 sequence. The calibration accuracy (unmagnified) is ±3% from 0.2 µsec/div to 2 sec/div; ±5% at 5 sec/div.VARIABLE Control
Uncalibrated variable time/div control. Multiplies the sweep time/div of any step of the TIME/DIV switch by a factor variable from 1 to >2.5.
POSITION Controls
Coarse horizontal positioning to position any portion of the trace to the center of the crt. The FINE control has a range of about 2 minor divisions.
MAGNIFIER Control
Provides direct readout magnification of the sweep (up to 50X). The magnifier also provides the 0.05 and 0.1 µsec/div sweep rates. Magnified sweep accuracy, ±5% (exception: 5 sec/div). Magnified sweep registration, ±1 minor division.
+ GATE OUT Connector
Provides a +20-volt (±15%) output pulse during sweep time.
Sweep Length
10.5 ±0.3 major division.
Sweep Amplifier (when used as an external horizontal amplifier)
HORIZONTAL VOLTS/DIV Switch
Calibrated volts/div steps of 0.2, 0.5, 1, 2, and 5 volts/div. Accuracy ±3% when calibrated to match the oscilloscope in which used. The amplifier is dc coupled to the EXT HORIZ IN connector.
VARIABLE Control
Multiplies volts/div of any step by a factor variable from 1 to >2.5.
EXT HORIZ IN Connector
Input R and C is typically 1 megΩ shunted by 40 pf.
Frequency Response
Dc to 400 kc. Response <30% down at 400 kc with the VARIABLE control set to CALIB position.
Maximum Input Voltage
±20 volts dc, or 20 volts peak ac.Construction Aluminum-alloy chassis with chrome-plated
side rails. Front panel is photo-etched and anodized.
- Peso netto
- 4.5 lb (4 lb 8 oz) / 2.043 kg
- Prezzo nel primo anno