Time Base 3B4
Tektronix; Portland, OR
- País
- Estados Unidos
- Fabricante / Marca
- Tektronix; Portland, OR
- Año
- 1964
- Categoría
- Aparato de medida y servicio (Equipo de laboratorio).
- Radiomuseum.org ID
- 231766
- Numero de valvulas
- 4
- Numero de transistores
- 20
- Semiconductores
- Gama de ondas
- - no hay
- Tensión de funcionamiento
- Alimentado desde otro equipo o aparato principal.
- Altavoz
- - - No hay salida de sonido.
- Material
- Metálico
- de Radiomuseum.org
- Modelo: Time Base 3B4 - Tektronix; Portland, OR
- Forma
- Chasis (tambien de autoradio)
- Ancho, altura, profundidad
- 4.25 x 6.5 x 14.5 inch / 108 x 165 x 368 mm
- Anotaciones
-
The Type 3B4 Time Base plug-in unit is designed to be used primarily with the Tektronix Type 561A Oscilloscope system; however, It can also be used with the other 560-Series systems that use 3B-Series plug-in units such as the Type 564 and Type 567. As part of an oscilloscope system, the Type 3B4 is used as an accurate time base generator or as a calibrated amplifier for externally generated deflection signals.
When used with the 567 or RM567, the 3B4 provides a time-base, but does not activate the digital readout circuitry. In the 565 or RM565, the 3B4 provides a vertical time-base for raster applications but does not provide retrace blanking.Sweep
TIME/DIV Switch
0.2 µsec/div to 5 sec/div in 23 steps, 1, 2, 5 sequence. The calibration accuracy (unmagnified) is ±3% from 0.2 µsec/div to 2 sec/div; ±5% at 5 sec/div.VARIABLE Control
Uncalibrated variable time/div control. Multiplies the sweep time/div of any step of the TIME/DIV switch by a factor variable from 1 to >2.5.
POSITION Controls
Coarse horizontal positioning to position any portion of the trace to the center of the crt. The FINE control has a range of about 2 minor divisions.
MAGNIFIER Control
Provides direct readout magnification of the sweep (up to 50X). The magnifier also provides the 0.05 and 0.1 µsec/div sweep rates. Magnified sweep accuracy, ±5% (exception: 5 sec/div). Magnified sweep registration, ±1 minor division.
+ GATE OUT Connector
Provides a +20-volt (±15%) output pulse during sweep time.
Sweep Length
10.5 ±0.3 major division.
Sweep Amplifier (when used as an external horizontal amplifier)
HORIZONTAL VOLTS/DIV Switch
Calibrated volts/div steps of 0.2, 0.5, 1, 2, and 5 volts/div. Accuracy ±3% when calibrated to match the oscilloscope in which used. The amplifier is dc coupled to the EXT HORIZ IN connector.
VARIABLE Control
Multiplies volts/div of any step by a factor variable from 1 to >2.5.
EXT HORIZ IN Connector
Input R and C is typically 1 megΩ shunted by 40 pf.
Frequency Response
Dc to 400 kc. Response <30% down at 400 kc with the VARIABLE control set to CALIB position.
Maximum Input Voltage
±20 volts dc, or 20 volts peak ac.Construction Aluminum-alloy chassis with chrome-plated
side rails. Front panel is photo-etched and anodized.
- Peso neto
- 4.5 lb (4 lb 8 oz) / 2.043 kg
- Precio durante el primer año
- 400.00 USD
- Procedencia de los datos
- -- Original-techn. papers.
- Autor
- Modelo creado por Pius Steiner. Ver en "Modificar Ficha" los participantes posteriores.
- Otros modelos
-
Donde encontrará 386 modelos, 374 con imágenes y 92 con esquemas.
Ir al listado general de Tektronix; Portland, OR
Colecciones
El modelo Time Base es parte de las colecciones de los siguientes miembros.