Oscilloscope 130C
Hewlett-Packard, (HP); Palo Alto, CA
- Pays
- Etats-Unis
- Fabricant / Marque
- Hewlett-Packard, (HP); Palo Alto, CA
- Année
- 1962–1966
- Catégorie
- Appareils de mesure et de dépannage (matériel de labo)
- Radiomuseum.org ID
- 218333
- No. de tubes
- 17
- No. de transistors
- 20
- Semi-conducteurs
- Gammes d'ondes
- - sans
- Tension / type courant
- Alimentation Courant Alternatif (CA) / 115; 230 Volt
- Haut-parleur
- - - Pas de sortie basse fréquence
- Matière
- Boitier métallique
- De Radiomuseum.org
- Modèle: Oscilloscope 130C - Hewlett-Packard, HP; Palo Alto
- Forme
- Modèle de table générique
- Dimensions (LHP)
- 480 x 425 x 180 mm / 18.9 x 16.7 x 7.1 inch
- Remarques
- The Hewlett Packard model 130C Oscilloscope is a versatile instrument for laboratory, production line, or industrial process measurements. Horizontal and vertical display sensitivity is 200 microvolts per centimeter and the measurement bandwidth is 500 kc. A sweep magnifier of up to X50 allows expansion of a trace to the equivalent of 500 centimeters for viewing waveform details.
SWEEP GENERATOR
INTERNAL SWEEP: 21 ranges, 1 µsec/cm to 5 sec/cm, accuracy within +/- 3%.
Vernier provides oontinuous adjustment between ranges and extends slowest sweep to at least 12.5 sec/cm.
MAGNIFICATION: X2, X5, X1O, X20, X50 overall sweep accuracy within +/- 5% for sweep rates which do not exceed a maxinwm rate of 0.2 µsec/cm.
AUTOMATIC TRIGGERING: Base line is displayed in the absence of an input signal.
Internal: 50 cps to 500 kc signal causing 0.5 cm or more vertical deflection and also from line voltage.
External: 50 cps to 500 kc, 0.5 volts peak-to peak or more.
Trigger Slope: Positive or negative slope of external sync signals or internal vertical deflection signals.
VERTICAL AND HORIZONTAL AMPLIFIERS
BANDWIDTH:
DC Coupled: DC to 500 kc
AC Coupled (input): 2 cps to 500 kc.
AC Coupled (amplifier): 25 cps to 500 kc at 0.2 mV/cm sensitivity.
Lower cut-off frequency (fco) is reduced as sensitivity is reduced; at 20 mv/cm fco is 0.25 cps. On less sensitive ranges, response extends to DC.
SENSITIVITY: 0.2 mV/cm to 20 V/cm. 16 ranges in 1,2,5,10 sequence with an attenuator accuracy within +/- 3%. Vernier permits continuous adjustment of sensitivity between ranges and extends mininwm sensitivity to at least 50 V/cm.
INPUT IMPEDANCE: 1 megohm shunted by 45 pf, constant on all sensitivity ranges.
RACK INSTALLATION.
The Model 130C is shipped from the factory ready for use as a bench instrument. The hardware necessary to rack-mount the instrument is packaged with the instrument.
90 W power consumption. Tubes listed once per type.
- Poids net
- 14 kg / 30 lb 13.4 oz (30.837 lb)
- Littérature
- - - Manufacturers Literature (Operating & Service Manual 1966)
- Auteur
- Modèle crée par Jean-Pierre Gérard. Voir les propositions de modification pour les contributeurs supplémentaires.
- D'autres Modèles
-
Vous pourrez trouver sous ce lien 393 modèles d'appareils, 361 avec des images et 138 avec des schémas.
Tous les appareils de Hewlett-Packard, (HP); Palo Alto, CA
Collections
Le modèle Oscilloscope fait partie des collections des membres suivants.