Oscilloscope 130C
Hewlett-Packard, (HP); Palo Alto, CA
- Produttore / Marca
- Hewlett-Packard, (HP); Palo Alto, CA
- Anno
- 1962–1966
- Categoria
- Strumento da laboratorio
- Radiomuseum.org ID
- 218333
- Numero di tubi
- 17
- Numero di transistor
- 20
- Semiconduttori
- Gamme d'onda
- - senza
- Tensioni di funzionamento
- Alimentazione a corrente alternata (CA) / 115; 230 Volt
- Altoparlante
- - - Nessuna uscita audio.
- Materiali
- Mobile di metallo
- Radiomuseum.org
- Modello: Oscilloscope 130C - Hewlett-Packard, HP; Palo Alto
- Forma
- Soprammobile con qualsiasi forma (non saputo).
- Dimensioni (LxAxP)
- 480 x 425 x 180 mm / 18.9 x 16.7 x 7.1 inch
- Annotazioni
- The Hewlett Packard model 130C Oscilloscope is a versatile instrument for laboratory, production line, or industrial process measurements. Horizontal and vertical display sensitivity is 200 microvolts per centimeter and the measurement bandwidth is 500 kc. A sweep magnifier of up to X50 allows expansion of a trace to the equivalent of 500 centimeters for viewing waveform details.
SWEEP GENERATOR
INTERNAL SWEEP: 21 ranges, 1 µsec/cm to 5 sec/cm, accuracy within +/- 3%.
Vernier provides oontinuous adjustment between ranges and extends slowest sweep to at least 12.5 sec/cm.
MAGNIFICATION: X2, X5, X1O, X20, X50 overall sweep accuracy within +/- 5% for sweep rates which do not exceed a maxinwm rate of 0.2 µsec/cm.
AUTOMATIC TRIGGERING: Base line is displayed in the absence of an input signal.
Internal: 50 cps to 500 kc signal causing 0.5 cm or more vertical deflection and also from line voltage.
External: 50 cps to 500 kc, 0.5 volts peak-to peak or more.
Trigger Slope: Positive or negative slope of external sync signals or internal vertical deflection signals.
VERTICAL AND HORIZONTAL AMPLIFIERS
BANDWIDTH:
DC Coupled: DC to 500 kc
AC Coupled (input): 2 cps to 500 kc.
AC Coupled (amplifier): 25 cps to 500 kc at 0.2 mV/cm sensitivity.
Lower cut-off frequency (fco) is reduced as sensitivity is reduced; at 20 mv/cm fco is 0.25 cps. On less sensitive ranges, response extends to DC.
SENSITIVITY: 0.2 mV/cm to 20 V/cm. 16 ranges in 1,2,5,10 sequence with an attenuator accuracy within +/- 3%. Vernier permits continuous adjustment of sensitivity between ranges and extends mininwm sensitivity to at least 50 V/cm.
INPUT IMPEDANCE: 1 megohm shunted by 45 pf, constant on all sensitivity ranges.
RACK INSTALLATION.
The Model 130C is shipped from the factory ready for use as a bench instrument. The hardware necessary to rack-mount the instrument is packaged with the instrument.
90 W power consumption. Tubes listed once per type.
- Peso netto
- 14 kg / 30 lb 13.4 oz (30.837 lb)
- Bibliografia
- - - Manufacturers Literature (Operating & Service Manual 1966)
- Autore
- Modello inviato da Jean-Pierre Gérard. Utilizzare "Proponi modifica" per inviare ulteriori dati.
- Altri modelli
-
In questo link sono elencati 393 modelli, di cui 361 con immagini e 138 con schemi.
Elenco delle radio e altri apparecchi della Hewlett-Packard, (HP); Palo Alto, CA
Collezioni
Il modello Oscilloscope fa parte delle collezioni dei seguenti membri.